频域OCT和HRT2在近视眼视盘参数测量的对比研究
作者:张铭志  文章来源:汕头大学?香港中文大学联合汕头国际眼科中心  点击数427  更新时间:2011/9/13  文章录入:毛进  责任编辑:毛进

Purpose To compare the optic disc topography measurements between the confocal scanning laser ophthalmoscopy (HRTII) and spectral-domain optical coherence tomography (Cirrus HD OCT) in normal subjects.
Methods A total of 105 eyes from 105 healthy subjects were included. The optic disc in each eye was imaged sequentially with HRT II (Heidelberg Engineering, GmbH, Dossenheim, Germany) and Cirrus HD OCT (Carl Zeiss Meditec, Dublin, CA). Paired t test was used to compare each topographic parameter between HRT2 and Cirrus HD OCT. The relationship between optic disc measurements of the 2 instruments was evaluated using a Pearson correlation analysis.
Results The mean age was 23±3.5 years. Optic disc topography measurements as determined by the 2 machines were significantly correlated (e.g disc area: r =0.78, p<0.001; rim area: r =0.45, p=0.011). HRT II measured significantly larger disc area and rim area than Cirrus HD OCT (disc area: 1.86±0.45 mm2 vs1.78±0.39 mm2, p=0.015; rim area: 1.50±0.29 mm2 vs1.34±0.27 mm2, p<0.01)
Conclusions There were significant differences in optic disc measurements as determined by HRT II and Cirrus HD OCT despite a significant correlation. Therefore, the two instruments should not be used interchangeably to obtain measurements of the optic disc for glaucoma diagnosis.